# Image-guided verification of daily radiotherapy set-up: imaging, matching and correction

Source: https://onco.cc/protocols/image-guided-radiotherapy-daily-setup-verification/  
OnCo record `image-guided-radiotherapy-daily-setup-verification` (Protocol). Data CC BY-NC 4.0, attribute "Data from OnCo (onco.cc)"; commercial use needs a licence.

## TL;DR

Before a radiotherapy treatment the radiographers often take a quick X-ray picture or scan on the treatment machine and compare it with the planning scan, to check that you are in the planned position. If the position is out, the couch is moved to correct it before the beam is switched on. How often this is done depends on the part of the body and the type of treatment.

## Summary

Image guidance compares an image taken on the treatment machine with the reference image from planning and corrects the difference. No source read grades its recommendations by evidence level; the fullest current document is the UK Radiotherapy Board's On target 2 (2021), read in full, beside the IAEA's implementation report (2019) and the AAPM practice guideline on IGRT systems (MPPG 2.b, 2021). The US ASTRO white paper update (2023) and ACR-ARS practice parameter (2025) could be read only as abstracts.

On target 2 asks each centre for site-specific protocols tailored to the site, with frequency, imaging dose and complexity reflecting treatment intent, site and fractionation; it says automatic matches should always be checked visually and gross errors always investigated. The IAEA has the radiation oncologist, with the multidisciplinary team, choose and prescribe the image guidance protocol, and says the protocol should state the frequency of imaging and the actions when a shift is above a threshold. MPPG 2.b, a guideline on the imaging systems, lists the duties of each profession. On imaging dose, On target 2 cites UK law (IR(ME)R) that exposures are justified and optimised, the IAEA says doses are justified, prescribed and reported, and MPPG 2.b says the dose must be measured at least annually.

They differ in three places. Correction rule: On target 2 recommends no action threshold for daily couch corrections where the correction is automated (every measured shift is applied), while the IAEA describes both that approach and action-level protocols that shift only above a set figure (its example is 2 mm), and MPPG 2.b asks physicists and oncologists to set repositioning thresholds. Online or offline: On target 2 notes a shift towards more online verification and says offline protocols are now much less commonly used, but that many of its site chapters still include them and that they so far have the largest evidence base; the IAEA says the choice depends on the site treated and the department's resources. Who decides: On target 2 puts online matching with trained radiographers at the machine with an independent second check; MPPG 2.b lists timely review of images among the radiation oncologist's duties; the IAEA gives the therapist final responsibility for the change, in some instances with input from the radiation oncologist, lists review and approval of online and offline images among the radiation oncologist's duties, and calls later offline review and approval good practice.

Imaging dose, each body's figures kept apart: On target 2 says three- and four-dimensional cone-beam CT dose is reported as approximately 1 to 5 cGy per scan and that daily imaging may result in total organ doses of 1 to 3% of the prescribed dose; the IAEA says kilovoltage cone-beam CT doses can be of the order of 1 cGy. On target 2 (5% of the prescribed dose) and the AAPM TG-180 abstract (5% of the therapeutic target dose) give the point above which imaging dose should be considered in planning. In the UK, national dose reference levels for cone-beam CT verification in adults were adopted in 2025, set by site as a weighted cone beam dose index, from 3.5 mGy (brain) to 21 mGy (prostate and gynaecological); this index is a different measure from the per-scan and organ doses above. A patient's own department's protocol comes first: this page explains the check and is not a set of instructions.

## Fields

- Kind: Protocol
- Last checked: 2026-11-17
- Also known as: IGRT; image-guided radiotherapy; image guidance; daily imaging before radiotherapy; treatment verification imaging; cone-beam CT before treatment; CBCT match; portal imaging; kV imaging; image matching; online correction; offline correction; no action level protocol; couch shift; set-up error; on-treatment verification

## Connected records

- protocols: [Bladder filling and rectal preparation for prostate radiotherapy](https://onco.cc/protocols/prostate-radiotherapy-bladder-and-rectal-preparation/), [Breast radiotherapy positioning, immobilisation and daily set-up verification](https://onco.cc/protocols/breast-radiotherapy-positioning-and-setup/), [Breathing motion in lung and upper abdominal radiotherapy: 4D CT, gating, breath-hold and compression](https://onco.cc/protocols/respiratory-motion-management-4d-ct-and-gating/), [Deep inspiration breath-hold for left-sided breast radiotherapy](https://onco.cc/protocols/breath-hold-left-breast-radiotherapy/), [Immobilisation mask for head and neck and brain radiotherapy, including mask anxiety](https://onco.cc/protocols/head-and-neck-radiotherapy-immobilisation-mask/), [Linear accelerator quality assurance: daily, weekly, monthly and annual machine checks](https://onco.cc/protocols/linear-accelerator-quality-assurance/), [Patient identification and pre-treatment time-out in radiotherapy](https://onco.cc/protocols/radiotherapy-patient-identification-and-time-out/), [Radiotherapy planning scan (CT simulation): positioning, skin marks and what the patient does](https://onco.cc/protocols/radiotherapy-planning-ct-simulation/), [Stereotactic radiosurgery and SBRT: planning and delivery safeguards](https://onco.cc/protocols/stereotactic-radiosurgery-and-sbrt-delivery-safeguards/)
- technologies: [Adaptive radiotherapy (online replanning)](https://onco.cc/technologies/adaptive-radiotherapy/), [IMRT / IGRT (modern external beam)](https://onco.cc/technologies/imrt-igrt/), [In-room imaging for radiotherapy (cone-beam CT, ExacTrac, CT-on-rails, HyperSight)](https://onco.cc/technologies/in-room-imaging-systems/), [Surface-guided radiotherapy (SGRT)](https://onco.cc/technologies/surface-guided-radiotherapy/)
- terms: [Fiducial markers](https://onco.cc/terms/fiducial-markers/), [GTV, CTV and PTV (target volumes)](https://onco.cc/terms/planning-target-volume/), [IMRT / IGRT / VMAT (modern external beam radiotherapy)](https://onco.cc/terms/imrt-term/), [Radiotherapy](https://onco.cc/terms/radiotherapy/)

---
JSON: https://onco.cc/api/v1/entities/image-guided-radiotherapy-daily-setup-verification.json