Dual-energy and spectral CT
CT scanned at two X-ray energies at once, so the scanner can tell iodine contrast from calcium and soft tissue, show exactly how much a tumour enhances, and remove the contrast digitally to save a second scan.
Overview
Because tissues absorb low- and high-energy X-rays differently, scanning at two energies lets software separate materials. Vendors reach this in different ways: two tube and detector pairs at different voltages (Siemens dual-source), one tube that switches voltage thousands of times a second (GE), a two-layer detector that sorts photons by energy (Philips), or a split filter or two sequential spins (Canon and others). The outputs are iodine maps that quantify tumour enhancement, virtual non-contrast images that replace a separate unenhanced scan, virtual monoenergetic images that boost contrast or suppress metal artefacts from hip implants and spinal hardware, and material-specific images such as virtual bone removal or calcium subtraction to reveal marrow disease.
In oncology it helps characterise kidney and adrenal masses, distinguish bland from tumour thrombus, detect hypervascular liver and pancreatic tumours at lower contrast doses, and see myeloma in the marrow. Photon-counting CT gives the same information natively; dual-energy remains the way most installed scanners obtain it.
- Rotating X-ray tube
- Lung nodule
How it works
Attenuation measured at two X-ray energy spectra is decomposed into material basis pairs (for example water and iodine), yielding iodine maps, virtual non-contrast and virtual monoenergetic images.
- Quantifies tumour enhancement and iodine uptake
- Virtual non-contrast saves a scan and dose
- Reduces metal artefact near implants
- Vendor-specific methods and protocols
- Some approaches lose temporal or spatial resolution
- Adds reading time and data
Latest papers
topQuery for this technology: (TITLE:"Dual-energy and spectral CT" OR ABSTRACT:"Dual-energy and spectral CT") AND (cancer OR tumor OR tumour OR oncology OR carcinoma OR lymphoma OR leukemia OR leukaemia OR myeloma OR sarcoma OR melanoma OR glioma). Results are unfiltered search hits about Dual-energy and spectral CT, not a curated reading list.
Similar pages
not linked directly; found by shared links- TechnologyX-ray radiography and fluoroscopy
Shares Canon Medical Systems, Philips, GE HealthCare, Siemens Healthineers and the tag machines-wave.
- TechnologyElastography and contrast-enhanced ultrasound
Shares Canon Medical Systems, Philips, GE HealthCare, Siemens Healthineers and the tag machines-wave.
- TechnologyMRI field strengths: 1.5 T, 3 T and 7 T
Shares Canon Medical Systems, Philips, GE HealthCare, Siemens Healthineers and the tag machines-wave.
- TechnologyIntraoperative MRI and CT
Shares Philips, GE HealthCare, Siemens Healthineers, CT (computed tomography) and the tag machines-wave.
- TechnologySPECT/CT
Shares GE HealthCare, Siemens Healthineers, CT (computed tomography) and the tag machines-wave.
- TechnologyMicrowave and radiofrequency ablation systems
Shares CT (computed tomography), Renal cell carcinoma, Hepatocellular carcinoma and the tag machines-wave.
- CompanyAngioDynamics
Shares Hepatocellular carcinoma, Pancreatic ductal adenocarcinoma and the tag machines-wave.
- CompanyViewRay (MRIdian)
Shares Hepatocellular carcinoma, Pancreatic ductal adenocarcinoma and the tag machines-wave.